IMPORTANT The Calibration Procedure that begins on
page 9 requires the sterilizer to be plugged in and
powered up while the adjustments are being per-
formed. Do not touch any components except for the
adjusting screws of the potentiometers. Use the Trim
Pot Adjustment Tool (RPI Part# RPT460) to set the
potentiometers.
DANGER: Failure to comply with
these instructions may result in an electrical shock,
which may result in severe personal injury or death.
5.0 - CALIBRATION PROCEDURE - PRESSURE/TEMPERATURE POTENTIOMETER ADJUSTMENTS